Incubator/Radiant Warmer Analyzer
Protect a baby’s safety by regularly performing all necessary preventive maintenance tests on infant incubators and radiant warmers. Life matters. Testing matters.
The equipment is an all-in-one Incubator/Radiant Warmer Analyzer that simplifies testing and ensures proper performance and safety of infant incubators, transport incubators and radiant warmers by complying to global standards (IEC60601-2-19, IEC 60601-2-20, IEC 60601-2-21).
Portable and intuitive-to-use, the equipment simultaneously tests temperature, airflow, sound, and humidity. Test results and parameters can be monitored in real-time on a large easy-to-read LCD screen, or be transferred to your computer wirelessly*.
Pass/Fail indicators are displayed at the end of tests for fast and easy troubleshooting, and a general testing option allows you to complete a quick test in just 15 minutes.
The equipment can stay in the incubator chamber up to 24 hours using its rechargeable battery, or for 48 hours on AC power, without compromising the integrity of the test environment.
- Description
- Additional information
Description
HIGHLIGHTS
Enables simultaneous measurement of humidity, airflow, sound, and temperature from six different points
Displays pass/fail indicators and real-time test results on a large LCD screen
Wireless* functionality supports fast and convenient downloading of test results and data
Able to choose from 10 different language interfaces
Portable and compact design, weighs less than three pounds
Has both Standards-based testing and a General Testing option which allows you to complete a test in as little as 15 minutes
Creates personalized test sequences for automatic test completion
Simplifies test set-up with color-coded temperature probes that match color of input ports
Complies with all global standards, including IEC 60601-2-19, IEC 60601-2-20 and IEC 60601-2-21
Measurements and tests specifications
Air convection peripheral temperature sensors for incubator (T1-T5) | 5 sensors |
Range: 0 °C to 50 °C | |
Accuracy: ±0.05 °C | |
Display resolution: 0.01 °C | |
Air convection temperature sensors for radiant warmers, sensors in pucks (black discs) | 5 pucks |
Range: 0 °C to 50 °C | |
Accuracy: ± 0.2 °C | |
Display resolution: 0.01 °C | |
Relative humidity | Range: 0 % to 100 %, |
Accuracy: +/- 3 % RH (0 % to 100 % non-condensing) | |
Display resolution: 0.1 % RH | |
Airflow | Range: 0.2 m/sec to 2.0 m/sec at 35°C, 50 % RH |
Accuracy: +0.1 m/sec | |
Display Resolution: 0.01 m/sec | |
Sound pressure | 30 dB(A) to 100 dB(A) |
Accuracy: ± 5 dB(A) | |
Display resolution: 0.1 dB(A) | |
IEC-61672-1 Class 2 from 31.5Hz to 8kHz | |
Surface temperature | -5 °C to 60 °C |
Accuracy: ± 0.5 °C | |
Display resolution: 0.05 °C | |
Skin temperature probe with reference thermometer | 0 °C to 50 °C |
Accuracy: ±0.05 °C | |
Display resolution: 0.01 °C |
Environmental specifications
Operating temperature | 10 °C to 40 °C |
Storage temperature | -20 °C to 60 °C |
Humidity | 10 % to 90 % non-condensing |
Altitude | 2000 m |
Ingress protection rating | IP-20 |
Safety | |
IEC 61010-1 | Overvoltage category none, pollution degree 2 |
IEC 61326-1 | Basic |
Electromagnetic compatibility (EMC)
Emissions classification IEC CISPR11: Group 1, Class A. | Group 1 has intentionally generated and/or use conductively coupled radio-frequency energy, which is necessary for the internal functioning of the equipment itself. Class A equipment is suitable for use in nondomestic locations and/or directly connected to a low voltage power supply network. |
USA (FCC) Intentional radiators | This device complies with part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) This device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation. (15.19) |
Korea (KCC) Class A Equipment (Industrial Broadcasting & Communication Equipment) | Class A: Equipment meets requirements for industrial electromagnetic wave equipment and the seller or user should take notice of it. This equipment is intended for use in business environments and not to be used in homes. |
Wireless module listing | |
FCC (United States) compliant (Class A) | FCC ID: X3ZBTMOD3 |
IC (Industry Canada) compliant | IC: 8828A-MOD3 |
CE (European) certified | CE0051 |
802.15.1 qualified | QD ID: B019224 |
Wireless radio | |
Frequency range | 2412 MHz to 2483 MHz |
Output power | 10 mW |
Interface | |
Push navigation buttons | Power on/off, test, select, back, and arrow keys |
User Preferences | Adjust backlight, display brightness, and set clock |
Recall and run templates on tester | General and personalized test templates allow for on-board test automation |
Templates | Time duration, frequency of data capture and tests |
Select user preferences | Units of measure, view test results of current and past tests on the Analyzer |
View battery life | Indicator bar shows life remaining |
Interface languages | English, French, German, Italian, Spanish, Brazilian Portuguese, Simplified Chinese, Japanese, Russian, Polish |
Size (HxWxL without sensors) | 23 cm x 21 cm x 6 cm (9.0 in x 8.5 in x 2.5 in) |
Total weight | 3.9 kg (8.5 lb) |
Weight with sensors | 1.4 kg (3 lb) |
Pucks (5) | 2.5 kg (5.5 lb) |
Carrying case | 1.1 kg (2.5 lb) |
Power Adapter – Universal voltage | Input: 100 V to 240 V with adaptors 50 Hz/60 Hz |
Output: 15V dc, 1.3 A maximum | |
Rechargeable lithium-ion battery, internal | 7.4 V, 7.8 Ah, 58 Wh 24 hour battery life with 30 second sample rate |
The Fluke Biomedical INCU II is an all-in-one Incubator/Radiant Warmer Analyzer that simplifies testing and ensures proper performance and safety of infant incubators, transport incubators and radiant warmers by complying to global standards (IEC60601-2-19, IEC 60601-2-20, IEC 60601-2-21).
Portable and intuitive-to-use, the INCU II simultaneously tests temperature, airflow, sound, and humidity.
4720996 INCU II-Cable K-Type Thermocouple
4721000 INCU II S Sensor Sound Sensor
4721017 INCU II A F Sensor Air Flow Sensor
4721021 INCU II H Sensor Humidity Sensor
4721039 INCU II R Sensor Red Temperature Sensor
4721042 INCU II G Sensor Green Temperature Sensor
4721056 INCU II Y Sensor Yellow Temperature Sensor
4721063 INCU II W Sensor White Temperature Sensor
4721074 INCU II B Sensor Blue Temperature Sensor
4721109 INCU II Spacers Tripod Spacers
4721111 INCU II R Puck Red Temperature Puck
4721127 INCU II G Puck Green Temperature Puck
4721130 INCU II Y Puck Yellow Temperature Puck
4721148 INCU II W Puck White Temperature Puck
4721153 INCU II B Puck Blue Temperature Puck
4721166 INCU II USB Type A to Micro B, USB Cable
4721194 INCU II Adapter AC Adapter
4715713 INCU II Pad Placement Pad Microfi ber
4715749 INCU II Carry Case Soft Carry Case
4724692 INCU II Puck Case Soft Carry Case Puck
4721175 INCU II ST Heater Skin Temperature Heater Assembly
Updating…
Ansur 3.1.4 Executive Software
Updating…
Additional information
Origin | USA |
---|---|
Type | Manual |