Rheoline MDR
Prescott’s Rheoline Moving Die Rheometer (MDR) has been designed to carry out, tried and proven test specifications in accordance with international standards. Providing fast, accurate data and meeting the requirements of a modern testing laboratory, the Rheoline MDR is ideally suited to both the production environment and for use in the development of elastomeric compounds.
Design conforms to ISO No. 6502 / ASTM D5289 Standards
- Description
- Additional information
Description
HIGHLIGHTS
Moving Die technology is a well-proven process that can significantly reduce your traditional testing and analysis times.
The digital electronics module and 16-bit PID algorithm apply quick, accurate temperature control with a variation of ± 0.03
At the heart of the Rheoline system is a rotor-less biconical die assembly with an internal cavity of 4.5 cm3. The lower die oscillates at 1.67 Hz over amplitudes of 0.5 °, 1.0 ° and 3.0 ° whilst the top die remains fixed and accurately measures the reaction torque, producing precise, repeatable data in an extensive range of testing environments.
Die Configuration | Rotor-less, Biconical, Sealed System |
Oscillation Frequency | 1.67 Hz |
Oscillation Amplitude | 0.5 ° (Std.), 1.0 ° and 3.0 ° |
Temperature Range | Ambient to 250 °C |
Temperature Control | 3 Term PID – Control Accuracy ± 0.03 °C |
Torque Range: | 0.001 to 250 dNm |
Real-Time Display | Standard torque curve Viscous curve Tan. Delta Cure Rate Programmable data points & limits Pass/Fail limits |
Data Calculations: | » Initial » Minimum » Maximum » Scorch Points » Percentage Cure Points » Reversion » Cure Rate » Delta Cure » Quality Control Pass/Fail |
Electrical | Single Phase 220/240V 50Hz / 110V 60Hz / 350VA |
Weight | 200 kg |
Dimensions:(W x D x H) | 575 mm x 570 mm x 1280 mm |
Pneumatics | Filtered Air, Min:0.41 Mpa / 60 psi / 4.14 Bar /4.2 kg/cm |
– A rheometer is a laboratory device used to measure the way in which a liquid, suspension or slurry flows in response to applied forces |
– It is used for those fluids which cannot be defined by a single value of viscosity and therefore require more parameters to be set and measured than is the case for a viscometer. |
Rheoline Auto-Load Fully-Automatic 16 sample loader Supported with: | The Auto-Load comes complete with three sample trays; each tray contains 16 sample slots. Samples are placed on the tray slots, one tray at a time is inserted into the loader, and the samples are then transferred automatically to the die cavity by way of a fabric belt. | |
The Advantages of the Auto-Test: – Up to 16 samples per a tray, queued ready for testing – Laboratory personnel free to perform other tasks – Automatic Pass/Fail of test results | ||
Compatible Auto-Test Instruments: – Rheoline Multi-Function Rheometer – Rheoline Moving Die Rheometer | ||
Technical Specification: | ||
Sample Capacity | 16 | |
Die Configuration | Sealed | |
Oscillation Frequency | 1.67 Hz | |
Oscillation Amplitude | 0.5, 1.0, 3.0 degrees | |
Temperature Range | Ambient to 250 ·c | |
Temperature Control | 3 term PIO, control accuracy +J- 0.03 °C |
Rheoline Auto-Test Semi-Automatic 4 sample tester Supported with: | The operator has the simple task of loading samples into the Auto-Test designated loading slots. The maximum number of samples loaded at one time is four, with the option to load an additional sample after each test (if required). | |
The Advantages of the Auto-Test: – Up to 4 samples queued ready for testing – Laboratory personnel free to perform other tasks – Automatic Pass/Fail of test results | ||
Compatible Auto-Test Instruments: – Rheoline Multi-Function Rheometer – Rheoline Moving Die Rheometer – Rheoline Mini MDR Lite | ||
Technical Specification: | ||
Sample Capacity | 4 | |
Die Configuration | Sealed 4.Scm3 | |
Oscillation Frequency | 1.67 Hz | |
Oscillation Amplitude | 0.5, 1.0, 3.0 degrees. | |
Temperature Range | Ambient to 250 ·c | |
Temperature Control | 3 term PIO, control accuracy+/- 0.03 °C |
Rheoline VSC Volumetric Sample Cutter | Our Rheoline Volumetric Sample Cutter can be relied upon to provide consistent samples of volume and size to be inserted into the die cavity. | |
This ergonomic designed pneumatic system enables precise user defined volume of uncured rubber and polymer material. The conveniently positioned and comfortable buttons are designed to work simultaneously to provide a safe working environment. | ||
Technical Specification: | ||
Sample Volume (adjustable): | 1 cm³ to 8 cm³ (normally 4.5 cm³) | |
Nominal Sample Diameter: | 34 mm | |
Nominal Sample Thickness @ 4.5 cm³: | 5.0 mm | |
Repeatability: | 0.25% | |
Delay between Squeeze and Cut (adjustable): | 1 sec to 10 sec | |
Pneumatics: | Filtered Air, Min: 0.5 MPa | 72 psi | 5 Bar | 5.09 kg/cm² | |
Weight: | 38 kg | |
Dimensions: (W x D x H) | 300 mm x 300 mm x 800 mm |
Option
Forced Cooling | An Air Cooling System to assist the operation of a Temperature Sweep. |
Pressure Module | This enables an additional pressure channel reading for Labline, when equipped with a pressure transducer. |
Software
The Multi-Function Rheometer Lite is supplied with the latest computer hardware, pre-installed with our most innovative data acquisition and analysis Labline Lite software for Microsoft Windows ® OS.
The Lite version licenses a maximum of 10 ‘Test Specifications’ with any number of test points. The software is also complete with a calibration module.
Key Features
Automatic Pass/Fail Result
Laboratory Statistics
Multilingual
Simple Instrument & Test Configuration
Pre-loaded Test Specifications
Produce Trend charts
Easy Result Recalculation
Report Printing Capabilities
Supports ODBC Database Connections:
– Microsoft Access ®
– Microsoft SQL Server ®
– IBM DB2 ®
Test Methods
Standard Cure (MDR mode, non isothermal)
Strain (Sweep LAOS)
Payne test
Stress Relaxation
Frequency Sweep
Conditioning
Linked tests any combination of test method and any duration
Channels
Elastic Torque
Viscous Torque
Tangent Delta
Cure Rate
Temperature (Upper)
Temperature (Lower)
Pressure in Cavity (optional upgrade)
Complex Torque
Loss Angle
Storage Modulus
Loss Modulus
Complex Modulus
Storage Compliance
Loss Compliance
Complex Compliance
Real Dynamic Visc.
Imag. Dynamic Visc.
Dynamic Complex Visc.
Additional information
Manufacturer | Prescott |
---|---|
Origin | USA |
Type | Semiautomatic |