Dry block temperature calibrator
The portable dry metal block temperature calibrator is temperature source and reference instrument in one device, for checking and calibrating of all kind of temperature instruments, such as thermometer, RTDs, Thermocouples, Temperature switches, etc.
- Description
- Additional information
Description
HIGHLIGHTS
Updating…
Temperature range: | +200°C…+1100°C |
Accuracy of indicated temperature: | ±3°C |
Resolution: | 0.01 / 0.1°C (selectable °C/°F/K) |
Stability of regulated temperature: | ±0.3°C at 1000°C |
Mean heating time: | 18°C per minute |
Mean cooling time: | 6°C per minute |
Tubular vertical furncace with an internal quartz pipe and interchangeable equalising block with 45 mm diameter Ceramic equalising block, 44 mm diameter x 175 mm depth
Insertion depth: | 220 mm (incl. upper insulation ) |
Counter-current forced air cooling system, enables to check even very short probes
Temperature uniformity: | at 1000°C: radial ±0.4°C (at 135 mm insertion depth) axial ±0,4°C (at 125 mm insertion depth) |
Interface: | RS232 |
Ramp function | |
Block with upper insulator: | 4 holes (dept 155 mm), diam. 7, 9, 11 and 13.5 mm |
Power supply: | 230 VAC (optional: 115 VAC), 50/60 Hz |
Power consumpon: | 850 VA |
Housing material: | metal |
Note: the stated accuracy is valid at ambient temperatur +10°C…+40°C and is guaranteed for one year.
Dimension: | 450 x 170 x 330 mm |
Weight: | 12 kgs net |
For mobile and stationary use
Microprocessor operated
Ramp function, thermostat tests
The ceramic equalizing block, whose large size suits the external dimensions, ensures a proper heat transmission as well as optimal stability and uniformity values. It’s interchangeability makes it possible to test sensors of various lengths, with diameters ranging from 1 mm up to 26 mm.
The device is equipped with a counter-current forced air cooling system, which keeps the temperature in the upper part of the well low. This feature allows checking also of very short probes, without heating the connecting head or the handgrip.
The calibrator runs on a voltage of 230 Vac (115V by request), 50/60Hz. |
A 2.5mt. cable with CEE polarized plug is supplied with the calibrator fitted with 2 conductors plus earth. |
This safety device with polarized plug is designed to reduce the risk of electric shock polarizing the phase of the power. |
Make sure that the plant is earthen correctly before switching the instrument on. |
Calibration and checking of all kind of thermometer
Calibration and checking of thermostats
Computer aided calibra?tion via RS232 interface
SOLAR-K-INS-01 SOLAR-I-INS-00 511.0.999.0007.0 | Ceramic adapter insert with 26 mm inner diameter to be used with Inconell insert with 25.5 mm outer diameter |
Inner insert made in Inconel 600, 25.5 mm outer diameter without drillings or millings, blank (for drilling/milling by customer) | |
Upper insulation without drillings | |
SOLAR-K-INS-01 SOLAR-I-INS-01 511.0.999.0007.0 | Ceramic adapter insert with 26 mm inner diameter to be used with Inconell insert with 25.5 mm outer diameter |
Inner insert made in Inconel 600, 25.5 mm outer diameter with 1 drilling (7…18 mm diameter) as per customer request | |
Upper insulation with suitable drilling | |
SOLAR-K-INS-01 SOLAR-I-INS-06 511.0.999.0007.0 | Ceramic adapter insert with 26 mm inner diameter to be used with Inconell insert with 25.5 mm outer diameter |
Inner insert made in Inconel 600, 25.5 mm outer diameter with 1 drilling (7 mm diameter) and 6 millings (2…6 mm diameter) as per customer request | |
Upper insulation with suitable drillings | |
SOLAR-K-INS-01 SOLAR-I-INS-06 511.0.999.0007.0 | Ceramic adapter insert with 26 mm inner diameter to be used with Inconell insert with 25.5 mm outer diameter |
Inner insert made in Inconel 600, 25.5 mm outer diameter with 1 drilling (7 mm diameter) and 12 millings (2…6 mm diameter) as per customer request | |
Upper insulation with suitable drillings | |
SOLAR-K-INS-04 511.0.999.0006.0 | Insert made in ceramic with 4 drillings: 7 – 9 – 11 – 13.5 mm (no adapter needed, outer diameter 42.5 mm) |
Upper insulation with suitable drillings | |
SOLAR-I-INS-04 511.0.999.0006.0 | Insert made in Inconel 600 with 4 drillings: 7 – 9 – 11 – 13.5 mm (no adapter needed, outer diameter 42.5 mm) |
Upper insulation with suitable drillings | |
SOLAR-I-INS-SB 511.0.999.0007.0 | Insert made in Inconel 600 without drillings (for drilling by customer) (no adapter needed, outer diameter 42.5 mm) |
Upper insulati?on without drillings (for drilling by customer) | |
INSERT-SOLAR 511.0.999.0007.0 | Black Body insert (no adapter required) , outer diameter 42.5 mm |
Suitable upper insulation |
AQ2sp software |
Temperature sensor calibration: |
Automatic calibration routine for all type of temperature sensors with advanced features such reference sensor stabilization, time-calibration data are automatically populated and stored for printing report. This procedure can also be ran in manual mode to allow the operator to manually enter the data. This is particular useful when calibrating temperature device such as digital thermometer or analogue thermometers. |
Stress test: |
Simply set MAX and MIN value and the software will run an automatic stress test on single sensor or a batch. |
Advance thermostat calibration: |
Simply set the open and close temperature limits, type of thermostat, typ of test and the software will automatically run the test. |
Additional information
Origin | Germany |
---|