Module of the portable electronic sphygmomanometer
A component module of the portable electronic sphygmomanometer . It provides NIBP (non‐invasive blood pressure) simulation supporting all major NIBP manufacturers.
- Description
- Additional information
Description
HIGHLIGHTS
Minimum Diastolic: 15 mmHg
Maximum Systolic: 275 mmHg
Amplitude Accuracy: better than 0.5%
Simulation type: oscillometric
– Pressure Units: | mmHg, mbar, kPa, inH2O, cmH2O |
– BP Simulation: | * Simulation type: oscillometric * Rate Range: 20 – 240 BPM * Accuracy: ±0.25 BPM in sync mode *±1 BPM otherwise * Amplitude: 0 – 2 mL * 1.25 mmHg into 500ml cuff * Amplitude Range: 0 – 150% * Amplitude Accuracy: better than 0.5% * AutoSettings: unlimited, user definable |
– Envelope Shift: | * ± 50 mmHg max * Minimum Diastolic: 15 mmHg * Maximum Systolic: 275 mmHg |
– Manometer: | *Pressure Range: 0.0 to 400.0 mmHg *Accuracy: ± 0.5 mmHg *Resolution: 0.1 mmHg – Regulated Pressure Source: *Pressure Range: 10.0 to 400.0 mmHg *Accuracy: ± 0.5 mmHg *Resolution: 0.1 mmHg |
– Leak Test: | *Automatic/manual Inflation *Automatic Timer *Leak Test Time: 30 – 600 seconds (user defined) *Target Pressure: 20 to 400 mmHg *Range: 0 to 200 mmHg/min *User Definable Presets: 12 *User Definable AutoSettings: unlimited |
– Power Supply:
+ External AC adapter
+ Internal rechargeable Li-Ion batteries (for 200+ simulations with full charge)
– Dimensions:
+ 98mm x 275mm x 97mm (3.87″” x 10.82″” x 3.80″”)
+ BP Unit (incl. A1 Base)
– Weight:
+ 1080g (2.38lb) BP Unit (incl. A1 Base)
+ 200g (0.44lb) wireless tablet interface
– Environment:
+ 15°C to 40°C, 10% to 90% RH,
+ Indoor Use Only, Category II”
vPad‐BP provides NIBP (non‐invasive blood pressure) simulation
Updating…
Updating
Additional information
Origin | Canada |
---|---|
Type | Semiautomatic |